Adjustment range
of output voltage
Min. voltage: 600 V~
Max. voltage: 5 kV~
Output current
Up to 10 mA
Adjustment range of internal chronometer Test
Time: It can be programmed up to 600 seconds.
Adjustment
range of current threshold Up to 10 mA
Voltmeter
accuracy ±
3% of measured value ±
0,05 kV
Test-voltage accuracy
Repeatability: ± 3%
of programmed value ±0,05 kV
(With main nominal supply voltage ± 1%)
Stability: ± 3%
of programmed value ± 0.05 kV
(With main nominal supply voltage ± 1%)
Milliammeter accuracy ±
5% of measured
value ± 0,1 mA
Chronometer
accuracy ±
0,5% of measured value ±
0,5 seconds
Current-threshold
accuracy ±
10 % of programmed value ±
0,1 mA
|
Dimensions 220
x 140 x 190 mm (cabinet outer parts)
Weight 3.5 kg
(including equipment, programming module, handbook and
terminals probes)
Supplied
attachments • A set of terminal probes used for
High Voltage • Power-supply cable with a tripole-jack
“NEMA” • Users' guide • Programming module PRM-100
Optional
accessories • EUED, protective
case.
Power supply Powered with
110 or 220 V~.
Operating
environmental conditions This equipment can only
be used in closed environments. Room temperature
can vary between 0 and 45 ºC, and humidity can
reach 90% RH without altering Hi-Pot performance. |