NOMINAL TEST VOLTAGES
250 V - 500 V - 1000 V.
Negative, DC voltage.
RESISTANCE MEASUREMENT RANGE
From 10 kΩ up to 200 GΩ.
SHORT CIRCUIT CURRENT
1.5 ± 0.3 mA.
TEST VOLTAGE ACCURACY
-0 / +15% for resistances between 10 MΩ and open circuit.
BASIC ACCURACY
R ≤ 50 GΩ: 5% of reading ± 2 digits.
R > 50 GΩ: 10% of reading ± 2 digits.
LOW RESISTANCE MEASUREMENT AND CONTINUIT TESTER
Range: 0.05 Ω to 150 Ω.
Accuracy: 5% of reading ± 2 digits.
Test current: up to 200 mA.
CContinuity indicator: Activates at R < 5 Ω ± 0.5 Ω.
BUILT-IN TIMER
5 - 10 - 15 - 30 seconds.
1 - 2 - 3 - 5 - 10 - 15 minutes.
PASS / FAIL TEST
1 - 2 - 5 - 10 - 20 - 50 - 100 - 200 - 500 MΩ.
1 - 2 - 5 - 10 - 20 - 50 - 100 GΩ.
VOLTMETER
True RMS.
Range: 0 - 600 V AC/DC.
AAccuracy: 3% of reading ± 2 digits.
OVERVOLTAGE PROTECTION
600 V.
ADDITIONAL FEATURES
• Automatic Polarization Index.
• Guard terminal.
• Hold function.
|
SAFETY CLASS
IEC 61010-1:1990, IEC 61010-1:1992 - 2.
E.M.C
In accordance with IEC 61326-1.
ELECTROSTATIC IMMUNITY
In accordance with IEC 1000-4-2.
ELECTRO MAGNETIC IRRADIATION IMMUNITY
In accordance with IEC 61000-4-3.
POWER SUPPLY
Built-in, NiMH battery. Autonomy: 1000 tests of 5 seconds with an interval of 25
seconds according IEC 61557-2. During the measurement is used a test resistance
wich the value is UN x (1000 Ω / V).
BATTERY CHARGER
Built-in, with external power supply of 9 V - 0.5 A.
OPERATION TEMPERATURE
-5°C to 50°C.
STORAGE TEMPERATURE
-10°C to 60°C.
HUMIDITY
95% RH (without condensation).
EQUIPMENT WEIGHT
Approx. 0.5 kg.
DIMENSIONS
102 x 195 x 46 mm.
INCLUDED ACCESSORIES
2 Measuring test leads.
1 GUARD test lead.
1 Power supply.
1 Carrying bag.
1 Operating instructions.
|