Test voltages (T.V.) 500,
1.000, 2.500, 5.000 V directly, one button selectable. 500 V to 5 kV in 100 V steps. DC, negative.
Maximum
resistance reading 5 TΩ @ 5kV.
Short
circuit current 1,5 ± 0,5 mA
Test voltages
accuracy ± 3% of nominal test voltages on R
>/= 10 GΩ
Megohmmeter basic accuracy ± 5% of
reading ± 3 digits (1 MΩ to 500 GΩ
at any test voltage)
Advanced
features Automated Polarization Index computing Automated
Dielectric Absorption Ratio computing Pass-fail test
/ Programmable timer Step voltage test Memory
for 4000 readings Real time clock and calendar
Built-in
printer Prints elapsed time, actual voltage and resistance
measured each 5 seconds.
Built-in
memory Stores up to 4000 insulation measurements.
Interface USB.
Built-in
chronometer Shows elapsed time in mm:ss format. Count
starts automatically for each measurement.
MegaLogg2 Software
Friendly, easy to use software. Tests are represented in graph and tabular
views. With automatic report generator, including the operator´s commentaries.
Environmental
protection IP54 with closed lid.
P 11012111
|
Safety class In accordance with IEC 61010-1
E.M.C In
accordance with IEC 61326-1
Electrostatic
immunity In accordance with IEC 61000-4-2
Electro
magnetic irradiation immunity In accordance with
IEC 61000-4-3
Power supply Internal
rechargeable 12 V - 2,3 Ah battery
Battery
charger For 100 - 240 V~ mains
Operating
temperature range -5°C to 50°C
Storage
temperature range -25°C to 65°C
Humidity range 95% RH (non
condensing)
Weight Approx. 3.6 kg
Dimensions 274
x 250 x 124 mm
Included accessories •
3 Measuring test leads 1,80 m. • 1 GUARD test lead,
1,80 m. • 1 Charger power cord.
• 1 USB Cable.
• 1 MegaLogg2 software.
• 1 User guide.
• 1 Carrying bag.
CE
mark
|