Test voltages (T.V.)
5 kV - 10 kV - 15 kV - 20 kV
Maximum
resistance reading 1 TΩ @ 5 kV (T.V.) 2 TΩ
@ 10 kV (T.V.) 3 TΩ @ 15 kV (T.V.) 4 TΩ
@ 20 kV (T.V.)
Short circuit current 500 µA
Test
voltages accuracy ± 2% of nominal test voltages on
R ³ 10 GΩ
Megohmmeter accuracy
class 2 (± 2% of full scale deflection)
Internal current
limiting resistance 10 MΩ @ 5 kV (T.V.) 20 MΩ
@ 10 kV (T.V.) 30 MΩ @ 15 kV (T.V.) 40 MΩ
@ 20 kV (T.V.)
Analogue indicator up to 99
mm scale length, taut band, with mirror
Environmental
protection IP54 with closed lid.
Safety class
In
accordance with IEC 61010-1
Electrostatic immunity
In
accordance with IEC 61000-4-2
Electro magnetic
irradiation immunity In accordance with IEC 61000-4-3
|
Power supply Internal
rechargeable 12 V - 7 Ah sealed lead
acid battery
Battery charger For 110-127
or 220 V~ mains supply
Operating temperature
range -5°C to 50°C
Storage temperature
range -25°C to 65°C
Humidity range
95%
RH (non condensing)
Altitude maximum 2000m
Weight
Approx.
9,8 kg
Dimensions 378
x 308 x 175mm
Included accessories •
Measuring test leads, 1,80 m. (2) • GUARD test lead,
1,80 m. • Protective case. • Charger power cord. •
Operating instructions
CE
mark
|