|
Four test voltages
(T.V.) 0,5 kV - 1 kV - 1,5 kV
- 2,5 kV
Maximum resistance reading 1 TΩ
@ 0,5 kV (T.V.) 2 TΩ @ 1 kV
(T.V.) 3 TΩ @ 1,5 kV (T.V.) 5 TΩ
@ 2,5 kV (T.V.)
Short circuit current
1 mA
± 2 %
Test voltages accuracy ±2% of
nominal test voltages on R ³ 1 GΩ
Megohmmeter
accuracy class 2 (± 2 % of full scale deflection)
Internal
current limiting resistance 500 KΩ @ 0,5 kV
(T.V.) 1 MΩ @ 1 kV (T.V.) 1,5 MΩ @ 1,5 kV (T.V.) 2,5 MΩ @ 2,5 kV (T.V.)
Analogue
indicator up to 99 mm scale length, taut band, with
mirror (thus avoiding parallax errors)
Environmental
protection IP54 with closed lid.
Safety class
In accordance
with IEC 61010-1
E.M.C In accordance with IEC 61326-1
and 61000-4-3
I11012711
|
Electrostatic immunity
In
accordance with IEC 61000-4-2
Power supply
Internal
rechargeable 12 V 2.3 Ah, sealed lead acid battery
Battery
charger 100-240 V~ mains supply
Operating
temperature range -5°C to 50°C
Storage
temperature range -25°C to 65°C
Humidity
range 95% RH (non condensing)
Altitude maximum 2000m
Weight
Approx.
3,9kg
Dimensions 274
x 250 x 124 mm
Included accessories •
Measuring test leads, 1,80 m. (2) • GUARD test lead,
1,80 m. • Charger power cord. • Protective case. • Operating instructions
CE
mark
|