TEST CURRENT
100 A, 200 A (True DC).
The current may be continuously adjustable from 0 to 100%.
RESISTANCE RANGES
0-1999 µΩ (test current: up to 200 A).
0-19.99 mΩ (test current: up to 100 A).
READING RESOLUTION
0.1 µΩ for R < 200 µΩ.
1 µΩ for R < 2000 µΩ.
10 µΩ for R < 20 mΩ.
OUTPUT VOLTAGE
3.5 Vdc @ 200 A.
4.5 Vdc @ 100 A.
5.8 Vdc (without load).
MAXIMUM LOAD RESISTANCE
10 mΩ @ 200 A.
30 mΩ @ 100 A.
MEASUREMENT PRINCIPLE
Four-terminal, U/I.
THERMAL PROTECTION
Protects all sensitive components, avoiding any damage due to overheating.
CONTINUOUS OPERATION TIME
At 200 A this equipment may be used continuously for aprox. 1 minute before the
thermal protection activates.
At 100 A this equipment may be used continuously
for aprox. 15 minutes before the thermal protection activates.
BASIC ACCURACY
± 0.5 % of reading ± 2 digits.
ADVANCED FEATURES
Digital direct reading of very low resistances in the alphanumerical display,
with up to 4½ digits.
Very fast and accurate measurements.
BUILT-IN MEMORY
It allows the storage of 200 tests readings.
|
SERIAL DATA OUTPUT
RS232 @ 4800 bps. For download the stored values.
ENVIRONMENTAL PROTECTION
IP54 (with closed lid).
SAFETY CLASS
Meets the requirements of IEC 61010-1:1990, IEC 61010‑1:1992 amendment 2.
POWER SUPPLY
Mains: 220-240 V~.
OPERATING TEMPERATURE RANGE
-5°C to 50°C.
STORAGE TEMPERATURE RANGE
-25°C to 65°C.
HUMIDITY RANGE
95 % RH (non condensing).
EQUIPMENT WEIGHT
Approx. 11 kg.
DIMENSIONS
502 x 394 x 190 mm.
INCLUDED ACCESSORIES
2 Combined current and potential leads - up to 15 m.
1 Power cord.
1 Ground cable.
1 RS232 cable.
1 Built-in thermal printer.
1 Case for the accessories.
1 User guide.
ACCESSORIES WEIGHT
Approx. 20.5 kg.
|