:: home > micro-ohmmeters > high current > max. current 200 A
MPK-204 - TECHNICAL SPECIFICATIONS

Test currents (T.C.)
1 mA, 10 mA, 100 mA, 1 A, 10 A, 200 A
Each current may be continuously adjustable from 0 to 100%

Resistance ranges
0-200 µΩ and 0-2000 µΩ @ 200 A(T.C.)
0-2000 µΩ and 0-20 mΩ @ 10 A(T.C.)
0-20 mΩ and 0-200 mΩ @ 1 A(T.C.)
0-200 mΩ and 0-2000 mΩ @ 100 mA(T.C.)
0-2000 mΩ and 0-20 Ω @ 10 mA(T.C.)
0-20 Ω and 0-200 Ω @ 1 mA(T.C.)
For each test current, ranges are automatically selected for optimal reading.

Resolution
0,1 µΩ @ 200 A(T.C.)
1 µΩ @ 10 A(T.C.)
10 µΩ @ 1 A(T.C.)
100 µΩ @ 100 mA(T.C.)
1 mΩ @ 10 mA(T.C.)
10 mΩ @ 1 mA(T.C.)

Measurement principle
Four-terminal, Kelvin-type.

Continuous operation time
At 200 A this equipment may be used continuously for aprox. 3 minutes before the thermal protection activates. At 10 A or less, there is not a limited time for continuous operation.

Thermal protection
Protects all sensitive components, avoiding any damage due to overheating.

Basic accuracy
± 0,25 % of reading ± 2 digits

Advanced features
Digital direct reading of very low resistances in the alphanumerical display, with up to 4½ digits.
Very fast and accurate measurements.

Test current measurement
Although the current is digitally measured, an analogue taut band instrument show the result  in order to facilitate the evaluation. This is specially useful when measuring inductive loads, so that the operator can verify easily when the test current has been stabilized

Serial data output
RS-232 @ 4800 bps. Suitable for data collection in an external serial printer, computer or data-logger.

Environmental protection
IP54 with closed lid.

Safety class
Meets the requirements of IEC 61010-1/1990, IEC 61010  1/1992 amendment 2

Power supply
Mains or internal battery powered
Battery is rechargeable, sealed lead-acid, 12 V 7 Ah
Mains: 220-240 V~
Internal battery is useful for up to 10 A .
Mains power is useful for any test current, including 200 A

Built-in battery charger
For 220-240 V~ mains

Operating temperature range
-5°C to 50°C

Storage temperature range
-25°C to 65°C

Humidity range
95% RH (non condensing)

Altitude maximum
3000m

Weight
Approx. 16 kg.

Dimensions
502 x 394 x 190mm

Included accessories
Combined current and potential leads for 200 A - 3,5m (2)
Combined current and potential leads for 10 A - 1,8m (2)
Aluminum case for the accessories
Power cord.
Operating instructions

 

 I04033001