Test currents (T.C.) 1 mA, 10 mA,
100 mA, 1 A, 5 A
Resistance
ranges 0-20 mΩ @ 5 A (T.C.) 0-600 mΩ
@ 5 A (T.C.) 0-200 mΩ @ 1 A (T.C.) 0-2000 mΩ @ 100 mA (T.C.) 0-20 Ω @ 10 mA (T.C.) 0-200 Ω @ 1 mA (T.C.)
Resolution 1 µΩ @ 5 A (T.C.)
Output voltage Up to
10 Vd.c. (open circuit) @ 1 A (T. C.)
Measurement
principle Four-terminal, Kelvin-type.
Basic
accuracy ± 0,25 % of reading ± 2 digits
Advanced
features Digital direct reading of very low resistances
in the alphanumerical display, with up to 4½ digits.
Very fast and accurate measurements.
Serial
data output RS-232 @ 4800 bps. Suitable for data
collection in an external serial printer, computer or
data-logger.
Environmental protection IP54
with closed lid.
Safety class Meets
the requirements of IEC 61010-1/1990, IEC 61010 1/1992
amendment 2
E.M.C In accordance with
IEC 61326-1
|
Electrostatic immunity In
accordance with IEC 1000-4-2
Electro magnetic
irradiation immunity In accordance with IEC 61000-4-3
Power supply Internal battery powered Battery
is rechargeable, sealed lead-acid, 12 V 7 Ah
Built-in
battery charger For 110-127 or 220-230 V~.
mains
Operating temperature range -5°C
to 50°C
Storage temperature range -25°C
to 65°C
Humidity range 95% RH (non
condensing)
Altitude maximum 2000m
Weight Approx.
8,7 kg
Dimensions 378 x 308 x
175 mm
Included accessories • Combined
current and potential leads with alligator clips 1,8m.
(2) • Charger Power cord. • Operating instructions •
RS-232 cable. • Sinthetic bag, for cables and instrument.
CE mark
|